TE-660 Dew point On-line / Transmitter

TE-660 Dew point  On-line / Transmitter

Description

TE-660 Dew point transmitter with a measuring range down to -60°C dp is ideal for applications in compressed air systems, plastic dryers and industrial drying processes. An integrated auto-calibration procedure permits a measurement accuracy of <2 °Cdp.

The measured values for dew point is available on an analog 4-20 mA. Integration into the measurement task is simplified by the compact design and the exceptionally robust stainless steel housing.

Installing
If installing directly into the process, a stop valve should be at both sides of the installation point. This makes it easy to remove the transmitter for maintenance and calibration.


Monitor for TE-660 On-line
If power supply and display are required, TK-100 On-line monitor is included in TK-660 On-line.

 

Key Features

  • Fully calibrated at TEKHNE factory
  • Innovated Polymer technology
  • Measuring range:-60 to +20 ̊C dp
  • Accuracy:± 2 ̊C dp (at 25 ̊C)
  • Pressure:up to 8 MPa
  • TE-660 On-line with monitor
  • TE-660 Transmitter without monitor
  • Fast response time

Specification

TE-660 Transmitter
Measurement range: -60 to 20 °Cdp
Accuracy: ±2 °Cdp (at 25 °C)
Response time t90: < 5 min (-20 °Cdp0-60 °Cdp)
                             < 15 sec (-60 °Cdp0-20 °Cdp)
Analog output: 4-20 mA(3-wire technology) RL < 500 Ohm
Resolution of analog output: 2 µA
Temperature dependence: ±5ppm of the measuring span / °C  (Deviating from 20 °C)
Power supply: 18 to 28 VDC
Power consumption at 24VDC: <20mA + load current
                                             with auto calibration: 100 mA + load current
Operating pressure: Up to 8 MPa
Electrical connection: M12 x1  5-pin plug
Sensor protection: Stainless steel sintered filter
Temperature/humidity operating range: -40 to 70 °C / 0 to 100 %RH
Storage temperature: -40 to 60 °C
Electromagnetic compatibility: EN61326-1 EN61326-2-3 Industrial environment
                                            FCC Part 15 ICES-003 ClassB


Monitor for TE-660 On-line
Display: 5 digit LED
Measurement Ranges: -60 to 20 °Cdp
Outputs: 4-20 mA
Alarm Relays: 3 relay contact
Operating environment: 0 to +50 °C (35 to 85 %RH)
Storage environment: -10 to +70 °C (max 60 %RH)
Power supply: 100 to 240 VAC, 50/60 Hz
Sensor cable: Cable, screened twisted pair (x2), 7/0.2 type 1, common drain.
                   (Max length 1,000 m)

 

Downloads

Brochures, Manual, Quick Start Guides, Software etc.